• School of Information Engineering and Automation, Kunming University of Science and Technology, Kunming 650500, China;
HEJianfeng, Email: 120112624@qq.com
Export PDF Favorites Scan Get Citation

In the process of positron emission tomography (PET) data acquiring, respiratory motion reduces the quality of PET imaging. In this paper, we present a correction method using three level grids B-spline elastic method to correct denoised and reorganized sinograms for respiratory motion correction. Using GATE simulates NCAT respiratory motion model to generate raw data which are used in experiment, the experiment results showed a significantly improved respiratory image with higher quality of PET, and the motion blur and structural information were fixed. The results proved the method of this paper would be effective for the elastic registration.

Citation: FENGShuo, CUIRui, HEJianfeng, MALei. Respiratory Motion Correction in Positron Emission Tomography Imaging Using Elastic Registration Based on Sinogram Data. Journal of Biomedical Engineering, 2014, 31(5): 1005-1010. doi: 10.7507/1001-5515.20140189 Copy

  • Previous Article

    Tumor Segmentation of Brain MRI with Adaptive Bandwidth Mean Shift
  • Next Article

    A Fast Iterative Reconstruction Algorithm of Sparse Angular CT Based on the Selective Total Variation