Export PDF Favorites Scan Get Citation

According to the working principle of vertical ultrasonic vial washing machine, receiving respective force of small water droplets on the inside wall of vials and the minimum air velocity of blowing off water droplets can be obtained based on the analysis of water-droplet-related parameters. The inside wall model of 7 mL vial created by GAMBIT was divided into fine grids. Then the Realizable k-epsilon Two Equation Turbulence Model was adopted and the flow field of vial by FLUENT software was simulated when air was flushing inside the wall. In that case, the optimal position, inner diameter and the corresponding minimum air velocity of needle can be acquired to meet the needs of vial washing machine applied to 7 mL vial.

Citation: ZHANGHaowei, LIZhen, LIUYing, LIUHaigang, PENGDelian, WEIGuoqin. Numerical Simulation and Optimization Research of Needle Parameters in Vial Washing Machine. Journal of Biomedical Engineering, 2014, 31(5): 1057-1060,1064. doi: 10.7507/1001-5515.20140198 Copy