• Department of Orthopaedics, Tianjin Dongli Hospital, Tianjin, 300203, P. R. China;
YINQingwei, Email: yqw1233068@sina.com
Export PDF Favorites Scan Get Citation

Objective To introduce a medical metal debris adsorption apparatus and to discuss the application value in removal of locking plate internal fixation. Methods Between January 2011 and May 2013, 83 cases underwent removal of locking plate fixation. There were 44 males and 39 females, aged 21-59 years (mean, 35.4 years). The time from the first operation to internal fixation removal was 16-40 months (mean, 22.6 months). Thirty-two screws of 9 patients were removed difficultly due to slide fastener. The screws were removed eventually after plates and screws were separated by grinding locking screw with manganese steel drill. In the process of grinding, the medical metal debris adsorption apparatus was used in 16 screws of 4 patients (group A), and was not used in 16 screws of 5 patients (group B). The operation time, removal time of single slipping screw, intraoperative blood loss, and the score of metal debris residues were compared between 2 groups. And the postoperative complications were recorded. Results A total of 658 screws were removed, and 32 slipping screws were found, accounting for 4.86%. The lock screw slipping rates of 3.5 mm and 4.5 or 5.0 mm respectively were 6.53% and 2.31%; and the lock screw slip ping rates of internal fixation time < 2 years and > 2 years respectively were 3.21% and 6.08%. The operation time, removal time of single slipping screw, intraoperative blood loss, and the score of metal debris residues of group A were significantly lower than those of group B (P < 0.05). Re-fracture (1 case) and incision infection (1 case) occurred in group B, and no complication was observed in group A. The cases of 2 groups were followed up 8-35 months (mean, 23.4 months). No heterotopic ossification occurred in both groups during the follow-up. Conclusion The medical metal debris adsorption apparatus can significantly shorten the operation time, reduce the residual metal debris and surgical bleeding when removing the slipping screws in removal of locking plate internal fixation.

Citation: YINQingwei, ZHANGHaibin. MANUFACTURE AND CLINICAL APPLICATION OF MEDICAL METAL DEBRIS ADSORPTION APPARATUS. Chinese Journal of Reparative and Reconstructive Surgery, 2014, 28(10): 1221-1224. doi: 10.7507/1002-1892.20140265 Copy

  • Previous Article

    BIOMECHANICAL STUDY ON REPAIRING DISTAL TIBIOFIBULAR SYNDESMOSIS INJURIES WITH ARTIFICIAL LIGAMENTS
  • Next Article

    ESTABLISHMENT OF MICROINVASIVE MODLE OF CHRONIC ROTATOR CUFF INJURY IN RATS