• Department of Cardiovascular Surgery, Xijing Hospital, the Fouth Military Medical University, Xi’an 710032, P.R.China;
Export PDF Favorites Scan Get Citation

Objective To design a novel stentless porcine aortic bioprosthesis and test the feasibility and its function in vitro after the valve was implanted by a modified method. Methods Six stentless porcine aortic bioprosthesis were divided into two groups according to different implantation, single layer suture group: new improvement stentless porcine aortic bioprosthesis sutured with single layer was implanted; double layer suture group: stentless porcine aortic bioprosthesis developmented by our laboratory used double layer suture was implanted. Each group contained three scales: 23 mm ,25 mm and 27 mm. Analogue ex vivo aortic valve replacement was performed , the feasibility of the new implantation was detected. Effective orifice area, transvalvular pressure gradient and regurgitation ratio were recorded at the cardiac output of 2.0 L/min, 3.5 L/min, 5.0 L/min and 7.0 L/min under the guideline of International Organization for tandardization (ISO)5840. Results The average aortic valve implantation time used for single layer suture and tradition double layer suture were 50 min and 70 min respectively. The transvalvular pressure gradient in the single layer suture group were significantly lower than those in double layer suture group under the flow of 5.0 L/min in 23 mm valve and 27 mm valve (13.51±0.51 mm Hg vs. 14.44±0.99 mm Hg, 7.36±0.19 mm Hg vs. 7.53±0.28 mm Hg;P<0.01);and the effective orifice area in the single layer suture group were larger than those in double layer suture group in the same case(1.87±0.06 cm2 vs. 1.76±0.08 cm2, 2.26±0.07 cm2 vs. 2.16±0.05 cm2;P<0.01). There was no statistically difference in other parameters between both groups. Conclusion The novel design of new improvement stentless porcine aortic bioprosthesis used single layer suture has good hemodynamic characteristics as the nature structure . The modified suture method decrease the implantation time.Nemerical data of the evaluation in vitro show that the difference between single layer suture group and double layer suture group in effective orifice area,transvalvular pressure gradient and regurgitation ratio have
no statistical significance. This experiment is the foundation of the animal and clinical experiment in the future.

Citation: XU Shu,YANG Jian,YI Dinghua,et al .. Rudiment Research and in Vitro Evaluation of a New Improvement Single Layer Suture Stentless Porcine Aortic Bioprosthesis. Chinese Journal of Clinical Thoracic and Cardiovascular Surgery, 2009, 16(2): 118-122. doi: Copy