Objective To describe the design and application of a new intramedullary controlled dynamic nailing(ICDN).〓〖WTHZ〗Methods The new ICDN was made of a stainlesssteel alloy, consists of a nail shaft, innerrod, tensile screw, end cape and two interlocked screws. From December 2005 toFebruary 2007, 48 cases of femoral shaft fractures were treated with ICDN. There were 16 females and 32 males with a mean age of 36.6 years (range, 16-48years). Injury was caused by traffic accident in 26 cases, falling from a height in14, sportsrelated in 7 and crush injury in 1 case. According to Winquist classification, Thirtyseven cases were classified as typeⅠ(13 cases of typeⅠ-1,23 cases of Ⅰ-2 and 1 case of Ⅰ-3), eleven as type Ⅱ fracture ( 2 cases oftype Ⅱ-1, 8 cases of Ⅱ-2 and 1 case of Ⅱ-3). The time from injury to admission was an hour to a week. Emergency operation was performed in 19 cases and the other 29 patients were given operation within 3 days. Results All cases were followed up for an average of 5 months (3-12 months). Among them, the average healing period was 21 weeks (826 weeks) and no nonunion, delayed union and breakage of nailings occurred. The complications included 2 infections, 5 pains, and 4 deformities. Based on Johner and Wrubs standard, the results were excellent in 39 cases, good in 7 cases and fair in 2 cases, The excellent and good rate was 95.8%. Conclusions ICDN has excellent biomechanical characteristics which can provide a flexible fixation in treating femoral shaft fractures. The operation is simple and is one of the good techniquesin treating femoral shaft fractures.
Citation: PAN Tao,WANG Guodong,WANF Jun,et al.. DEVELOPMENT AND CLINICAL APPLICATION OF AN INTRAMEDULLARY CONTROLLED DYNAMIC NAI LING. Chinese Journal of Reparative and Reconstructive Surgery, 2007, 21(12): 1362-1365. doi: Copy