• Department of Orthopaedic Surgery, Shajing People’s Hospital, Shenzhen Guangdong,518104,P.R. China;
Export PDF Favorites Scan Get Citation

Objective To ascertain whether augmentation pedicle screw fixation with polymethylmethacrylate (PMMA) can enhance the stability of unstable thoracolumbar burst fractures of osteoporotic spine. Methods Six fresh frozen female osteoporotic spines (T10-L5) were harvested and an anterior and posterior columnunstable model of L1 was made. Each specimen was fixated with plate and the stability test were performed by flexion, extension, axial rotation and lateral bending. The test of fatigue was done with MTS 858.The tests were repeated after screws were augmented with PMMA. To compare the biomechanical stability of 6 different conditions:○anormal specimens(control), ○bdefectmodel fixed with plate, not augmented and not fatigued, ○cafter fatigued, not augmented, ○dscrews augmented with PMMA, not fatigued, ○e after augmented and fatigued. ResultsIn ○b,○d and ○e conditions, the ranges of motion(ROM) were 6.23±1.56,4.49±1.00,4.46±1.83 inflexion and 6.60±1.80,4.41±0.82,4.46±1.83 in extension. There was no significant difference (P gt;0.05), they were significantly smaller than those in ○a and ○c conditions (8.75±1.88,1.47±2.25 and 8.92±2.97,12.24±3.08) (P lt;0.01).Conclusion The results demonstrated that augmentation pedicle screws fixation with PMMA can increase the stability of osteoporotic spine.

Citation: FAN Shicai,LIU Shixue,DENG Yuexing,et al.. AN IN VITRO BIOMECHANICAL EVALUATION OF EFFECT OF AUGMENTATION PEDICLE SCREW FIXATION WITH POLYMETHYLMETHACRYLATE ON OSTEOPOROTIC SPINE STABILITY. Chinese Journal of Reparative and Reconstructive Surgery, 2004, 18(3): 168-170. doi: Copy

Copyright © the editorial department of Chinese Journal of Reparative and Reconstructive Surgery of West China Medical Publisher. All rights reserved