west china medical publishers
Keyword
  • Title
  • Author
  • Keyword
  • Abstract
Advance search
Advance search

Search

find Keyword "降主动脉mechanical ventilation" 1 results
  • The Impact of Mechanical Ventilation on Pulse Oximetry in Thoracic Cavity

    摘要:目的: 评价机械通气对胸腔内脉搏氧饱和度的影响。 方法 :以食道、气管和降主动脉作为胸腔内脉搏氧饱和度的监测位点,将改制后的氧饱和度探头分别固定于上述部位,并连接于同一监护仪上。纯氧通气,待上述氧饱和度容积波波形和读数稳定,停止机械通气30s。以录像的方式记录机械通气停止前后30 s内食道、气管和降主动脉SpO2容积波和读数的变化。同时记录舌SpO2。 结果 :机械通气时,食道、气管和降主动脉三个监测位点均可获得异常高大的SpO2容积波;停止通气时,异常高大的氧饱和度波形消失。食道、气管和降主动脉脉搏容积波变异率分别为112%,74%,302%。降主动脉脉搏容积波的变异率明显高于食道和气管( 〖WTBX〗P <005)。机械通气停止前后30s内食道、气管和降主动脉的SpO2读数变化无显著差异(〖WTBX〗P >005)。 结论 :机械通气对胸腔内食道、气管和降主动脉氧饱和度读数无影响,主要影响是脉搏容积波。且各位点间脉搏氧容积波受呼吸的影响不同。Abstract: Objective: To investigate the impact of mechanical ventilation on pulse oximetry in thoracic cavity. Methods : After dogs anesthesia induction and thoracotomy, pulse oximeters were simultaneously placed at esophagus, trachea, and descending aorta, and connected with the same monitor for SpO2 monitoring. During ventilation with 100% oxygen, the mechanical ventilator was temporarily switched off for 30 seconds after high quality PPG waveforms and SpO2 readings were obtained. SpO2 signals and readings from esophagus (SeO2), trachea (StraO2), descending aorta (SDAO2) shown on the monitoring screen were recorded by the SONY video before and after stopventilation. And StonO2 were also recorded. Results : Abnormally largeamplitude PPG waves were found in normal waves at monitoring sites of esophagus, trachea, and descending aorta in all animals during ventilation; however, they disappeared without ventilator. The variation rate in ventilationinduced PPG amplitude were 112%, 74%, 302% at esophagus, trachea and descending aorta respectively. The PPG amplitude variation rate from SDAO2 was higher than that from SeO2 and StraO2 (〖WTBX〗P <005). However, the SpO2 readings obtained from pulse oximetries in all sites were no significantly statistical difference within 30s before and after temporarily stopventilation (〖WTBX〗P >005). Conclusion : Abnormally amplitude PPG waveforms from oximetry probe placed at esophagus, trachea, and descending aorta were induced by ventilation. The Variation rate in ventilationinduced PPG amplitude was various at different monitoring sites. The SpO2 readings from esophagus, trachea, and descending aorta were not significantly contaminated by ventilation.

    Release date:2016-09-08 10:12 Export PDF Favorites Scan
1 pages Previous 1 Next

Format

Content