  • Comparison of the Precision of Cone Beam CT and Electronic Portal Image Device in Guiding the Radiation Therapy for Thorax Tumor

    目的 分析比较锥形束CT(CBCT)与MOSAIQ两系统用于测量胸部放射治疗(放疗)摆位误差情况。 方法 2011年2月-9月,随机选择21例用热塑模固定的胸部调强治疗患者,进行同次摆位的CBCT图像采集和电子射野影像(EPID)正侧位图像采集。将CBCT图像与定位CT图像进行匹配,图像经iViewGT系统传输至MOSAIQ系统图像处理模块中与计划生成的数字重建图像(DRR)正侧位片进行匹配,分别记录两套系统图像匹配结果在X(左右)方向、Y(头脚)方向、Z(前后)方向的平移误差及CBCT的旋转误差值,计算其均值、标准差,进行统计学处理。 结果 本研究共获取图像106组,每组内包含一次CBCT图像和EPID图像。CBCT和MOSAIQ系统下EPID匹配结果平移误差在X、Y、Z方向分别为(−0.103 ± 0.240)cm、(−0.086 ± 0.342)cm、(0.017 ± 0184)cm和(0.005 ± 0.214)cm、(−0.004 ± 0.0.315)cm、(−0.113 ± 0.239)cm;旋转误差在X、Y、Z方向分别为(0.792 ± 1.173)°、(−0.130 ± 1.407)°、(0.793 ± 0.960)°,其中≤3°的概率分别为98.2%、97.2%、99.1%,最大绝对值分别为5°、3.4°、3.3°。两套系统在X、Y、Z三个方向的平移误差经配对样本检验,在Y方向P>0.05,在X和Z方向上P值均<0.05, 两系统的测量差值有直线负相关关系。 结论 CBCT和EPID在测量胸部放疗靶区中心的平移误差中差异有统计学意义,因此建议在做胸部放疗时有CBCT的情况下应优先使用。

    Release date:2016-09-08 09:11 Export PDF Favorites Scan
  • Rotation Errors Due to Neck Deformation in Nasopharyngeal Cancer Radiotherapy with Image Guiding

    【摘要】 目的 利用不同匹配区域对锥形束CT(CBCT)与定位CT(FBCT)分别配准,测量出鼻咽癌放射治疗中颈部的变形误差。 方法 分析2007年4月-2008年12月收治鼻咽癌患者23例,调整治疗床前198次CBCT扫描。将鼻咽部扫描CBCT图像匹配区域分为上下两个区域进行对比分析。其中上匹配区域为:上界为蝶窦上缘,下界为颈4下缘,侧界包括下颌骨外轮廓,前界为上颌窦1/2,后界为平棘突后缘;下匹配区域为:上界约颈4下缘,下界约胸2-3下缘,侧界包括椎体外轮廓,前界包括皮肤,后界平棘突后缘。匹配方式选择骨,比较匹配结果差异。 结果 选择上与下匹配区域结果除Y(头脚)方向旋转误差无统计学差异外,余均有统计学差异(Plt;0.05) 。差值在X(左右)、Z(前后)、Y(头脚)方向平移分别为(1.14±2.80)、(0.47±1.41)、(0.58±3.88) mm,旋转误差X、Y、Z方向分别为(0.90±1.98)、(0.80±2.03)、(0.68±1.90)°。 结论 鼻咽癌放射治疗中颈部区域存在一定变形误差,通过CBCT引导发现变形误差并进行正确纠正是必须的,结合临床实际及靶区与危及器官的变化为重新计划提供依据。【Abstract】 Objective To investigate the rotation errors due to neck deformation in nasopharyngeal cancer (NPC) radiotherapy with different match areas to register conebeam CT(CBCT) from image guiding and fanbeam (FBCT) from simulation. Methods A total of 198 pre-correction CBCT data sets from 23 NPC patients from April 2007 to December 2008 were retrospectively analyzed. The matching areas in CBCT images were divided into up and down region of interest (ROI). For the up ROI, the superior, inferior, left and right, anterior, and posterior boundary were set parallel with sphenoid sinus up side, C4 down side, mandible outside, and 1/2 of maxillary air sinus and acanthi. For the down ROI, the lines were set parallel with C4 down side, T2-3 down side, neck outside, skin surface and acanthi respectively in all directions. All registrations were performed automatically by bony anatomy and the results were compared. Results The registration results by the up and the down ROI showed significant difference except Y direction for rotation. The translation error was (1.14±2.80),(0.47±1.41),and (0.58±3.88) mm, respectively; and the rotation error was (0.90±1.98),(0.80±2.03),and (0.68±1.90) ° in X, Y, and Z direction, respectively. 〖WTHZ〗Conclusions〖WTBZ〗There are some significant deformation errors at neck areas in NPC radiotherapy. It is important to find out the deformation and correct it with CBCT image guiding. This kind of error information may provide clues for re-planning in addition to clinical practice and the changes of clinical targets and involved organs.

    Release date:2016-08-26 02:21 Export PDF Favorites Scan