west china medical publishers
Author
  • Title
  • Author
  • Keyword
  • Abstract
Advance search
Advance search

Search

find Author "JIANGXiao-qin" 1 results
  • Application of Liner-accelerator Two-degrade Collimator in the Treatment of Nasopharyngeal Carcinoma with Volumetric Modulated Arc Therapy

    ObjectiveCompare the two-degrade collimator (MLC) angle selection's impact on plan quality and operational efficiency for volumetric intensity-modulated radiotherapy (VMAT) in the treatment planning system, and to explore the scheme for treatment plan optimization. MethodsTwenty patients with nasopharyngeal carcinoma underwent the treatment between March and December 2013 were randomly selected and planned for SIBVMAT treatment with different parameters set in the range of 0-60°with 15°interval for collimator angles. Planned dose distribution to the target volumes, organs at risk, and monitor units were compared. ResultsAs the MLC angle increased, target conformal index and homogeneity index had a trend to became deteriorated. The optimal plans were 0°and 15°, while 45°and 60°plans gave poor protection for the organ at risk compare to other angle plans and the monitor units were significantly increased. ConclusionChange the MLC angle had visible impact on treatment plans,there was a trend to deteriorate with the MLC angle increased, but small changes in MLC angle range can theoretically reduce the influence from leakage radiation on the human body.

    Release date: Export PDF Favorites Scan
1 pages Previous 1 Next

Format

Content