This study was aimed to estimate the effect of different ProDisc-C arthroplasty designs after it was implanted to C5-C6 cervicalspine. Finite element (FE) model of intact C5-C6 segments including the vertebrae and disc was developed and validated. Ball-and-socket artificial disc prosthesis model (ProDisc-C, Synthes) was implanted into the validated FE model and the curvature of the ProDisc-C prosthesis was varied. All models were loaded with compressed force 74 N and the pure moment of 1.8 Nm along flexion-extension and bilateral bending and axial torsion separately. The results indicated that the variation in the curvature of ball and socket configuration would influence the range of motion in flexion/extension, while there were not apparently differences under other conditions of loads. The method increasing the curvature will solve the stress concentration of the polyethylene, but it will also bring adverse outcomes, such as facet joint force increasing and ligament tension increasing. Therefore, the design of artificial discs should be considered comprehensively to reserve the range of motion as well as to avoid the adverse problems, so as not to affect the long-term clinical results.
We observed the effect of vibration parameters on lumbar spine under different vibration conditions using finite element analysis method in our laboratory. In this study, the CT-images of L1-L5 segments were obtained. All images were used to develop 3D geometrical model using the Mimics10.01 (Materialise, Belgium). Then it was modified using Geomagic Studio12.0 (Raindrop Geomagic Inc. USA). Finite element (FE) mesh model was generated by Hypermesh11.0 (Altair Engineering, Inc. USA) and Abaqus. Abaqus was used to calculate the stress distribution of L1-L5 under different vibration conditions. It was found that in a vibration cycle, tensile stress was occurred on lumbar vertebra mainly. Stress distributed evenly and stress concentration occurred on the left rear side of the upper endplate. The stress had no obvious changes under different frequencies, but the stress was higher when amplitude was greater. In conclusion, frequency and amplitude parameters have little effect on the stress distribution in vertebra. The stress magnitude is positively correlated with the amplitude.