  • System Design of a Medical Constnt Temperature Flushing Pump

    The flushing pump which is applied to clean operative wound has no temperature controlling function up to now, and doctors have to prepare the flushing fluid that has previously been warmed. The flushing pump system with medical constant temperature designed in our laboratory can absorb flushing fluid at the room temperature, and then eject flushing fluid with the temperature in accordance with the requirements of operations at a controlled constant flow rate. The system combines flow rate control with temperature control functions. The flushing pump system includes flushing part, temperature controlling part, key inputting part, liquid crystal displaying part and exceptional situation monitoring part. The present paper introduces the design method and principle of each part of the system at first, and then gives the debug method of all the system parameters. Finally the paper discusses the performance of the system according to the result of the experiment.