摘要:目的:分析微创外科室间隔缺损(ventricular septal defect,VSD)封堵失败原因,以期提高术前超声心动图筛查水平。方法:回顾性分析25例微创外科VSD封堵失败改行修补术病例,对比超声表现及手术所见,归纳总结产生并发症的原因。结果:残余分流与VSD假性膜部瘤右室面具有多个出口和低估VSD大小密切相关;VSD合并主动脉瓣右冠瓣脱垂是主动脉瓣反流的主要原因;封堵器移位与低估VSD大小且使用偏心封堵器有关;原有三尖瓣反流加重和发生Ⅲ度房室传导阻滞VSD均位于隔瓣下方;封堵失败组较封堵成功组缺损偏大,差异具有统计学意义(Plt;0.05)。结论:超声心动图对VSD及其毗邻结构的细致评估,有助于严格适应证,提高手术成功率。 Abstract: Objective:To analyze the failure of perventricular closure of ventricular septal defect (VSD), in order to improve the preoperative echocardiography examination. Methods: Twentyfive cases underwent surgical repair after failure of perventricular closure of VSD were included in this study. With combination of echocardiographic and surgical findings, retrospective analysis of the failure of perventricular closure of VSD were attempted to summarize the cause of complications.Results: Residual ventricular communication was due to underestimation of size of VSD and pseudomembranous aneurysm resulting in multiple outlets of VSD on the right ventricle side; preoperative prolapse of rightcoronary cusp was the main reason for mild or greater than mild aortic valve regurgitation after eccentric device closure of VSD; Underestimation of the size of VSD and using eccentric occluder device were responsible for the displacement of VSD occluder device. Postoperative aggravated tricuspid regurgitation and Ⅲ°atrialventricular block (AVB) were attributed to VSDs located under the septal leaflet of tricuspid valve. The size of VSD in group of failed perventricular device closure of VSD was larger than that in group of successful device closure of VSD,and the difference was significant(Plt;0.05). Conclusion: Echocardiography vividly reveals VSD and adjacent structures, which should be used in accessing the anomaly and defect and formulating surgical plans to reduce surgical morbidity and mortality.