Objective To evaluate the application value of scanning laser angiography with a wide-field contact lens system in the diagnosis of choroidal melanoma. Methods Twenty-four patients with choroidal melanoma were randomly divided into two groups, who underwent fundus fluorescein angiography and indocyanine green angiography scanning with the wide-field contact and non-contact lens system respectively in order to acquire the 150 deg;wide-field and 30 deg;view image data. The quality of the images was comprehensively evaluated. Results Satisfying images were acquired from all of the 24 patients. Widefield contact lens system indicated the accurate adjacent relation between the lesion position and the other dissection mechanisms, and also provided the general information about the size of the tumor and the perfusion of fluorescien or indocyanine green in the blood vessels. At the same time, it enlarged the view scope 3-5 times, which make for the screening of the peripheral lesions. Conclusions Scanning laser angiography with a wide-field contact lens system has important application value in the diagnosis of choroidal melanoma. (Chin J Ocul Fundus Dis, 2006, 22: 166-169)